Study on Thermal Performance Evaluation and Standards Compliance of NeutralElectrodes

LI Jianing1DING Yi² WANG Pengfei³ ZHANG Chao³* (1.Centre Testing International Group Co.,Ltd.;2.ProSteri Testing Shanghai Co.,Ltd.; 3. KingPo Technology Development Co., Ltd.)

Abstract:With the widespreadapplicationof high-frequency electrosurgical devices inclinicalsurgery,thethermal performance and standards compliance of neutral electrodes,as critical safety components,are directly related to patient safety.BasedonthestandardGB97o6.202-2021,Medicalelectricalequipment—Part2-2:Particularrequirementsforthe basicsafetyandessentialperformanceof highfrequencysurgicalequipmentandhighfrequencysurgicalaccesories,this papersystematically interprets the heating mechanisms and standards compliance of neutral electrodes,and analyzes the scientificrationaleandimplementationrequirementsofthetestingcriteriaspecifiedinthestandard.Ithighlights theroot causes of discrepancies between the standard safety threshold (temperature rise ≤6∘C )and the clinical theoretical limit (temperature rise ≤12∘C ,evaluatestherationalityof graded test current levels( 350Ω-700mA )validatedbythethermal factormodel Θ=I2×t) ,and explores the risk characteristics ofhigh-current modes (>30A2S) . It aims to provide anin-depth interpretationof thecorerequirementsof the standard clauses,offercriticalguidance forneutral electrodemanufacturers inachieving standards compliance,optimize product designpathways,andensure the safetyand compliancein clinical application,thereby providing technical assurance for the safe implementation of high-frequency electrosurgical procedures.

Keywords: neutral electrode; thermal performance; high-current mode

0 引言

高频电外科设备是一种通过高频电流(通常200kHz~5MHz )作用于生物组织实现切割与凝血功能的医疗器械。(剩余5273字)

目录
monitor