可吸收性外科缝线中环氧乙烷残留量测定的不确定度评定

打开文本图片集
Uncertainty Evaluation of Detection of Ethylene-oxide Content in Absorbable Surgical Sutures
JIANG Chen-yang LIU Shu-chang (Weihai Institute for Food and Drug Control)
Abstract: In order to improve the accuracy of the detection results of ethylene-oxide content in medical equipment, absorbable surgical suture is taken as an example, and the uncertainty is evaluated. The ethylene-oxide content in absorbable surgical sutures is tested according to GB/T 14233.1-2022, Test methods for infusion, transfusion, injection equipment for medical use—Part 1: Chemical analysis methods. The uncertainty is analyzed and evaluated. At the 95% confidence interval, the expanded uncertainty is 4.420mg/Kg , and the range of tested results is (14.765±4.420)mg/Kg. and the established method can provide reference for the uncertainty evaluation of ethylene-oxide content testing in other medical equipment.
Keywords: absorbable surgical sutures, ethylene-oxide, uncertainty
0 引 言
环氧乙烷(EO)残留量测定项目是评定医疗器械产品是否安全有效的重要指标之一[1]。(剩余3509字)