微波消解-石墨炉原子吸收光谱法测定榆林小杂粮中的镉含量

打开文本图片集
Determination of Cadmium in Yulin Minor Cereals by Microwave Digestion- Graphite Furnace Atomic Absorption Spectrometry
SHAO Huahua, JIA Li, GAO Wenrui, ZHAO Huiming (Yulin Food Inspection and Testing Center, Yulin 719ooo, China)
Abstract: In this paper, an analytical method for determining cadmium content in minor cereals by microwave digestion-graphite furnace atomic absorption spectroscopy was established.The results of method validation showed that the linear relationship between cadmium in the range of 0μgL-1 to 2.5μgL-1 was good, the correlation coefficient r=0.999 4 ,the relative standard deviation of the method precision was 1.7% to 3.1% , the detection limit was (204号 0.00074mg⋅kg-1 ,and the recovery rate was 95.5% to 102.0% ,all of which met the requirements of trace detection. A total of30 kinds of minor cereals in5categories were tested,andthecadmium content in all samples was lower than the requirements of the GB 2762—2022 limit standard.
Keywords: cadmium; microwave digestion; graphite furnace atomic absorption spectrometry; minor cereals
榆林小杂粮不仅是当地农民收人的主要来源之一,也是助力乡村经济发展的龙头产业[1-2]。(剩余3651字)