医用泄漏电流分析仪校准装置研究

打开文本图片集
【摘要】为了解决医用漏电分析仪电流示值误差参数的溯源问题,研制一种医用漏电分析仪校准装置。本文介绍了该校准装置的研究方案和硬件实现,接着通过一系列实验验证了该校准装置的性能,最终确定医用漏电分析仪电流示值误差参数溯源方法。
【关键词】医用漏电分析仪;渥尔曼电路;Howland电流泵
【DOI编码】10.3969/j.issn.1674-4977.2024.01.070
Research on Calibration Methods for Medical Leakage Current Tester
SHANG Shuangshuang
(Liaoning Institute of Measurement, Shenyang 110004, China)
Abstract: To solve the traceability issue of the current indication error parameters of the medical leakage current tester, a calibration device for the medical leakage current tester has been developed. This article introduces the research plan and hardware implementation of the calibration device. Then, the performance of the calibration device is verified through a series of experiments, and finally, the traceability method for the current indication error parameters of the medical leakage current tester is determined.
Keywords: medical leakage current tester; Wollmann circuit; Howland current pump
1概述
我国早在2007年就已出台泄漏电流分析仪或类似设备的检定规程JJG 843—2007《泄漏电流测试仪》,但专用的检定装置鲜有报道。(剩余6275字)