【摘要】文章依据JJG 692—2010《无创自动测量血压计检定规程》并结合工作经验,分析了电子血压计的外观、静态压力、示值重复性等检定项目,重点讨论了静态压力的测量方法,并对其结果进行了不确定度评定。

【关键词】电子血压计;静态压力;不确定评定

【DOI编码】10.3969/j.issn.1674-4977.2024.01.069

Analysis and Discussion on Metrological Verification Items of Non-invasive Automatic Measurement Electronic Sphygmomanometer

CHEN Shurong

(Beijing Xicheng Institute of Metrology, Beijing 100055, China)

Abstract: Based on JJG 692—2010 Non-invasive Automated Sphygmomanometers and daily work experience, the article analyzes the verification items of electronic blood pressure meters, such as appearance, static pressure, and repeatability of indication. The focus is on discussing the measurement method of static pressure, and evaluating the uncertainty of the results.

Keywords: digital electronic sphygmomanometer; static pressur; uncertainty evaluation

目前,用于医疗卫生的电子血压计已经被列入《实施强制管理的计量器具目录》中。(剩余3835字)