无创呼吸机测量参数示值误差的测量不确定度评定

  • 打印
  • 收藏
收藏成功


打开文本图片集

【摘要】无创呼吸机是一种以非创伤方式通过口鼻为使用者提供持续的正压气流,用于控制或辅助使用者呼吸的机械通气设备。它是一种应用广泛,并且非常重要的医疗仪器,其工作参数的准确性关乎患者的救治效果和生命安全。本文对无创呼吸机的测量参数示值误差进行详细的测量不确定度评定分析。

【关键词】无创呼吸机;不确定度;示值误差

【DOI编码】10.3969/j.issn.1674-4977.2024.05.078

Assessment of Measurement Uncertainty for Indicated Value Error of NonInvasive Ventilators’ Measurement Parameters

LI Dunyong

(Guangxi ZhuangAutonomous Region Institute of Metrology &Test, Nanning 530007, China)

Abstract: Non-invasive ventilators are mechanical ventilation devices that provide continuous positive airway pressure to the user through the mouth and nose in a non-invasive manner, controlling or assisting the user’s breathing. They are widely used and extremely important medical instruments, where the accuracy of their operational parameters is crucial for the therapeutic effect and the safety of patients’ lives. This article conducts a detailed analysis of the measurement uncertainty assessment for the indicated value error of the measurement parameters of non-invasive ventilators.

Keywords: non-invasive ventilator;uncertainty;indicated value error

1概述

无创呼吸机又称气道正压通气呼吸机,是一种以非创伤方式通过口鼻为使用者提供持续的正压气流,用于控制或辅助使用者呼吸的机械通气设备,其包括但不限于睡眠呼吸暂停治疗设备、家用呼吸支持设备、家用呼吸机。(剩余2798字)