• 注册
  • 礼品卡激活

上消化道镍钛合金支架移位后的回收和重置

  【摘要】 目的 探讨如何安全有效的取出或重置移位的上消化镍钛合金支架。方法 本组采用自行研制的上消化镍钛合金支架取出器共进行了17例,18例次移位的上消化镍钛合金支架的回收或重置。结果 本组使用自行研制上消化镍钛合金支架取出器共重置食管上段支架4例次,取出支架2例次;重置中段支架3例次;重置食管下段支架5例次,取出支架2例次;重置十二指肠支架2例次。所有上消化镍钛合金支架的取出或重置操作均未出现上消化道大出血、穿孔等严重并发症。结论 采用自行研制的上消化镍钛合金支架取出器对移位的上消化镍钛合金支架进行取出或重置是安全和有效的。

  【关键词】支架 食管狭窄 介入放射学

  Retrieve/Reset the Displaced Nickel-Titanium Alligation Stent in Upper Gastrointestinal Tract Department of Interventional Radiology. Wei ning,Li Guojun,Zu Maoheng,et al.Affiliated Hospital of Xuzhou Medical College.Xuzhou,221002,China.

  【Abstract】Objective To discuss how to retrieve/reset the displaced Nickel-Titanium alligation stent in upper gastrointestinal tract.Methods In this serial, the displaced upper gastrointestinal tract Nickel-Titanium alligation stent was retivere/reset in 17 cases, 18 times.Results A self produced Nickel-Titanium alligation stent reset instrument was used 4 times in displaced upper esophagus stent resetting and 2 times retrieving; It was used 3 times in displaced middle esophagus stent resetting, It was used 5 times in displaced middle esophagus stent resetting and 2 times retrieving, It also was used 2 times in displaced duodenum stent resetting. All practice did not lead to bleeding or perforation et al.Conclusion The self produced Nickel-Titanium alligation stent reset instrument is safe and efficient in retrieving/resetting the displaced Nickel-Titanium alligation stent.

   【Key Words】Stent;Esophageal stenosis;Interventional radiology

  【中图分类号】R57 【文献标识码】A 【文章编号】1005-0019(2008)11-0004-03

  上消化道狭窄是较为常见的临床表现,造成这种临床症状的常见原因包括食管癌、食管化学性烧伤、食管手术和十二指肠梗阻性疾病等[1]。(剩余1457字)

龙源期刊网

收录3000种正版杂志,种类遍及时政、财经、文学、生活、娱乐、教育、学术等诸多门类,并同时以互联网和无线方式发行。

全国免费客服电话:400-6565-456

版权所有© 1997-2012 龙源期刊网

互联网出版许可证:新出网证(京)字066号

京公海网安备110108901919

电信与信息服务业务经营许可证:京ICP证060024   

关注龙源官方微博,热点期刊精彩分享: